Examples of Equipment/Reagent Quality
Control
Immunohematology Review 484
Blood storage refrigerators & freezers,
platelet incubators
Temperature recorder
Alarms
Waterbaths
Heat blocks
Centrifuges
Cell washers
Pipettes
Antisera
Reagent cells
Antihuman globulin
Not all inclusive. Follow manufacturer’s guidelines & lab’s Standard Operating Procedures Manual.
System for continuous temp monitoring & audible alarm.
Compare against thermometer daily. Calibrate as necessary.
Check high & low temp of activation quarterly.
Check temp daily.
Check temp daily. Periodically check each well.
Determine optimum speed & time for different procedures upon receipt, after repairs,
& periodically. Check timer every 3 mo, RPM every 6 mo (with tachometer).
Check tube fill level daily, AHG volume monthly. Verify time & speed quarterly.
Calibrate quarterly.
Test with pos & neg controls each day of use. Use heterozygous cells for pos controls.
Check for hemolysis. Test each day of use with pos & neg controls.
Check anti-IgG activity each day of use by testing Rh-pos cells sensitized with anti-D.