effects can be minimized by selecting the initial (low) tilt
value to correspond to a well-defined detent position if
the mechanical stage is so designed, such as a physical
stop at 0° tilt. With a properly maintained mechanical
stage, the uncertainty in the tilt angle difference Δθ is

estimated to be approximately 2 % for Δθ = 50 , with the
relative uncertainty increasing for smaller values of Δθ.


  1. Considering all of these sources of uncertainty, the
    measurement should be assigned an overall uncertainty
    of ±5 % relative.


a

b

. Fig. 6.20 a Stereo pair of a
machined screw thread—SEM/E–
T(positive) images; E 0 = 20 keV. b
Stereo pair with superimposed
axes for measurement of coordi-
nates needed for quantitative ste-
reomicroscopy calculations


6.5 · Making Measurements on Surfaces With Arbitrary Topography: Stereomicroscopy