636 Chapter 11. Dosimetry and Radiation Protection After irradiation the material is etched with a suitable chemical, such as ho ...
11.4. Active Dosimetry 637 A.1 FreeinAirIonChamberDosimetry As the name suggests, this type of ionization chamber measures expos ...
638 Chapter 11. Dosimetry and Radiation Protection Measurement Volume Air Enclosure Cathode Anode Photons Compensating Volume Co ...
11.4. Active Dosimetry 639 where kd is the correction due to variation in density of air. kh is the correction for the dependenc ...
640 Chapter 11. Dosimetry and Radiation Protection Wall (Cathode) Insulator HV Signal Anode Insulator Insulator Air Air Figure 1 ...
11.4. Active Dosimetry 641 This is all good as long as we wanted to measure the doseinside the chamber. How we relate this dose ...
642 Chapter 11. Dosimetry and Radiation Protection withμm.,enbeing the mass energy absorption coefficient. This coefficient shou ...
11.4. Active Dosimetry 643 Table 11.4.1: Typical uncertainties in air filled ionization chambers used in dosime- try. Source Unc ...
644 Chapter 11. Dosimetry and Radiation Protection p+(Source) p+(Drain) Silicon Substrate Al Al SiO 2 SiO 2 Oxide Polysilicon Si ...
11.4. Active Dosimetry 645 MOSFET dosimeters also have some disadvantages, some of which are listed below. Sensitivity to Bias ...
646 Chapter 11. Dosimetry and Radiation Protection The disadvantages of diamond dosimeters include the following. Dose Rate Dep ...
11.4. Active Dosimetry 647 11.4.DQuartzFiberElectroscope.................... Quartz fiber electroscope is a pocket dosimeter tha ...
648 Chapter 11. Dosimetry and Radiation Protection (^0200400) ++++++ + +++ ++ +++ ++ + (b) (^0200400) + + + + + ++++ + + + + + − ...
11.5. Microdosimetry 649 11.5Microdosimetry.............................. Microdosimetry is a rapidly evolving technique that is ...
650 Chapter 11. Dosimetry and Radiation Protection whereD(L) represents the normalized dose distribution as a function ofLET.Iti ...
11.5. Microdosimetry 651 whereD(y) represents the normalized dose distribution as a function of lineal energy. It can be evaluat ...
652 Chapter 11. Dosimetry and Radiation Protection equivalence can be approached by assuring that the energy losses of charged p ...
11.5. Microdosimetry 653 probability of this happening is small due to smaller volume (see Fig.11.5.1(a)). The dose measured in ...
654 Chapter 11. Dosimetry and Radiation Protection (b) (a) (d) (c) TEPC Tissue Wall Figure 11.5.1: Processes leading to measurem ...
11.5. Microdosimetry 655 B.3 Silicon Microdosimeter Microdosimeters based on silicon are a relatively newer development. As we w ...
«